Cecomp CS
Ranges and Resolution
See table below for ranges
Default engineering unit: psi
User configurable for other engineering units
Resolution is fixed for each engineering unit
Accuracy
Accuracy includes linearity, hysteresis, repeatability
Sensor hysteresis: ±0.015% FS, included in accuracy
Sensor repeatability: ±0.01% FS, included in accuracy
Pressure and absolute ranges
±0.1% full scale ±1 least significant digit
Vacuum and vacuum/pressure ranges
±0.25% full scale ±1 least significant digit
Display
3 readings per second nominal display update rate
4 digit LCD, 0.5"H and 5 character 0.25"H alphanumeric
White LED backlight active for 1 minute with button keypress
Batteries, Battery Life, Low Battery Indication
2 AA alkaline included
Approx. 150-1500 hours depending on backlight usage
Low battery symbol on display
Controls, Functions
Front button powers gauge on/off, activates backlighting for
1 minute, zeros gauge reference gauges, and cycles through
min./max. functions
Internal buttons for selection of engineering units and auto
shutoff time. Passcode protected calibration, min/max setup.
Min/Max Memory
Minimum and/or maximum readings stored in memory
Readings cleared or stored at shutoff
User configurable